Note 12: Resistance values greater than 180 k ohms cannot be measured by the OM-USB-TEMP in the
thermistor mode. The 180 k ohm resistance limit includes the total resistance across the current
excitation (±Ix) pins, which is the sum of the thermistor resistance and the lead resistances.
Note 13: For accurate three wire compensation, the individual lead resistances connected to the ±Ix pins
must be of equal value.
Throughput rate
Table 9. Throughput rate specifications
Number of Input Channels Maximum Throughput
1 2 Samples/second
2 2 S/s on each channel, 4 S/s total
3 2 S/s on each channel, 6 S/s total
4 2 S/s on each channel, 8 S/s total
5 2 S/s on each channel, 10 S/s total
6 2 S/s on each channel, 12 S/s total
7 2 S/s on each channel, 14 S/s total
8 2 S/s on each channel, 16 S/s total
Note 14: The analog inputs are configured to run continuously. Each channel is sampled twice per second.
The maximum latency between when a sample is acquired and the temperature data is provided
by the USB unit is approximately 0.5 seconds.
Digital input/output
Table 10. Digital input/output specifications
Digital type CMOS
Number of I/O 8 (DIO0 through DIO7)
Configuration Independently configured for input or output.
Power on reset is input mode.
Pull-up/pull-down configuration All pins pulled up to +5 V via 47 K resistors (default). Pull-down to ground
(GND) also available.
Digital I/O transfer rate (software paced)
Digital input – 50 port reads or single bit reads per second typ.
Digital output – 100 port writes or single bit writes per second typ.
Input high voltage 2.0 V min., 5.5 V absolute max.
Input low voltage 0.8 V max., –0.5 V absolute min.
Output low voltage (IOL = 2.5 mA) 0.7 V max.
Output high voltage (IOH = –2.5 mA) 3.8 V min.
Note 15: All ground pins on the OM-USB-TEMP (pins 9, 19, 28, 38) are common and are isolated from
earth ground. If a connection is made to earth ground when using digital I/O and conductive
thermocouples, the thermocouples are no longer isolated. In this case, thermocouples must not be
connected to any conductive surfaces that may be referenced to earth ground.
Memory
Table 11. Memory specifications
EEPROM 1,024 bytes isolated micro reserved for sensor configuration
256 bytes USB micro for external application use
Microcontroller
Table 12. Microcontroller specifications
Type Two high-performance 8-bit RISC microcontrollers