Owner's manual

3
1.3.1. Data Displays
The LCD displays the data in a continuous cycle, alternating in duration according to the
order shown in the table below:
Time on Screen
(seconds)
Data Reading Example of Reading on
Screen
8 Temperature (in
Celsius or Fahrenheit)
25.5 ˚C
2 Low alarm duration *
AL-L
1.55H
2 High alarm duration *
AL-H
1.05H
2 Minimum temperature
reading
MIN
23.2 ˚C
2 Maximum temperature
reading
MAX
30.1 ˚C
2 Low battery warning * BATT LO
* If relevant (only displayed in event of alarm/low battery)
Current Temperature Reading
This is the last temperature reading recorded by the MicroLite. If the sampling interval is
once per second, then the reading will be updated for each of the four seconds it
displayed on the LCD. Therefore you will be looking at the actual temperature recorded
at that very moment.
MicroLite Alarm
MicroLite displays an alarm notification whenever any alarm level is exceeded.
AL-L – The temperature reading is lower than its low alarm level (as defined in the
logger Setup).
AL-H – The temperature reading is higher than its high alarm level (as defined in the
logger Setup).
The alarm notification remains until the next time you download data to a PC.
To learn how to set alarm levels, refer to page 9.
The alarm duration on the logger’s LCD is calculated in hours. For example, the shortest
possible duration of 0.01 hours means the logger has been in alarm for between 0 and
36 seconds.
A duration of 1.55 hours means the logger has been in alarm for 93 minutes.
If the MicroLite is in a Low and High alarm simultaneously, then it will show the duration
for both alarms.
Note: When the MicroLite is in Cyclic mode, the Low or High alarm level displayed
represents the lowest or highest readings recorded since the MicroLite starting logging
data, irrelevant of how many cycles the logger has recorded.