Owner's manual
12
2. Click Change Password to open the Change Password dialog box.
3. Enter the current password in the Current Password field.
4. Enter the new password in the New Password field.
5. Enter the new password a second time in the Confirm New Password field to
confirm your new password.
6. Click OK.
Note: The password must include at least 4 characters and is case sensitive.
Temperature Calibration
1. Connect the MicroLite to the PC.
2. Click Logger on the main menu, and then click Calibration.
3. Enter the calibration password, and then click OK.
4. The Calibration dialog will open.
Figure 5: Calibration dialog box
5. Enter the Reference value: this is the temperature value you expect the logger to
display.
6. Enter the MicroLite value: this is the temperature value the logger is currently
displaying, when compared with the reference value.
7. Click Calibrate.
To restore default calibration, click Default. This will remove any calibration applied to
the MicroLite.
2.3.6. Communication Setup
Communication between the PC and the MicroLite unit takes place automatically
whenever you send a command to the MicroLite. However, the Communication Setup
dialog box can be used for more advanced communication options.
Click Logger on the main menu and select Com setup to open the Communication
Setup dialog box: