Owner's manual
Temperature/Pressure Adapter
6
recording mode. This will display the starting
time of the recording. The display will show the
starting time of the last recording and allow the user to enter a new starting
time.
The
STATUS
button will advance the item that is flashing. Press and hold the
STATUS
button to scroll this item. Press the
RECORD
button to change to
the next item which will then start to flash. This will set the starting time of the
recording which will appear on the display or printout after a recording has
been made. Note that the previous starting time is always displayed when a new
recording is started.
When the Service Logger is in the recording mode, it will prevent the user from
changing the settings. The user will only be allowed to check the recording rate
when they press the
MENU
ENTER
button.
Press the
MENU
ENTER
button to return to the operating mode and begin the re-
cording.
If option 1 or 3 is selected, an "R" will be dis-
played in the upper right hand corner of the
display. An "R" indicates a recording is in
progress.
While you are recording it is possible to check
how much memory has been used. Press and
hold the
STATUS
button until the display
shows the percent of available remaining memory and the starting time of the
recording.
As the recording continues this number will decrease. The Service Logger will
continue to record until the recording is terminated manually or when all of the
available memory has been used. When 0% Memory Left is displayed, this will
indicate that almost all of the available memory has been used and the Service
Logger will shortly terminate the recording automatically.
Sample of a typical printout at the beginning of recording:
The time is specified using the format D(ays):H(ours):M(inutes):S(econds).
This convention is used in the printout and the display.
Service Logger
Temp (F) Pressure PSI Time (D:H:M:S)
TEMP PRESSURE R
76°F 0 PSI
MON:10:35:0
56% MEM. LEFT