Owner's manual
Temperature/Pressure Adapter
2
Temperature sensor is not responding.
This is the Operating Mode display for the
OM-60-MOD-TP adapter. Temperature and
pressure are continuously measured and
displayed. This is the mode the Service Logger
will go to after it is turned on.
CHANGING SETTINGS
Every function of the Service Logger can be set through a series of menus. Set-
tings are stored in memory and remain there even without the main and backup
power source until altered by the user. Each sensor will have settings unique to
its function and the Service Logger will select and display the correct menus for
each sensor that is plugged in.
Changing Recording Time and Sampling Rate
Press and hold the
MENU
ENTER
button until the
display changes to the Recording Rate menu.
The user may select the sampling rate and
recording time for the Service Logger. Use
the
STATUS
or
RECORD
buttons to change
the recording time and sampling rate according
to the table listed below.
Recording Time Sampling Rate
40 Days 15 Minutes
2 Weeks 5 Minutes
60 Hours 1 Minute
30 Hours 30 Seconds
15 Hours 15 Seconds
5 Hours 5 Seconds
When the
MENU
ENTER
button is pressed 4 more times (to cycle through the re-
maining functions), the recording time and the sampling rate will be saved into
memory and the Service Logger will return to the operating mode. Otherwise
the unit will return to the operating mode after approximately 1 minute and
changes will be saved into memory.
TEMP PRESSURE
**** 0 PSI
TEMP PRESSURE
74°F 0 PSI
RECORDING RATE
15HR EVERY 15 SEC