Owner's manual
28
4.3.5 Configuration (continued)
F) Input Contact Closure
Contact 1
: Disable, Normally Open, Normally Close
Contact 2
: Disable, Normally Open, Normally Close
Disabled: Contact is dysfunctional.
Normally Open: Contact is open. If the contact is closed, there will be an Output High
or Low, depending on which output is selected. Read Sensor page will display:
Normal.
Normally Close: Contact is closed. If the contact is open, there will be an Output High
or Low, depending on which output is selected. Read Sensor page will display: Active.
Active Contact Output
: Active High, Active Low. The Output can be High or Low when
the status on the input contact is changed.
Output Type
: Unlatch, Latch, Clear Latch
G) Terminal Server
TCP/UDP*
: The iSE supports TCP and UDP protocols (default is TCP). If UDP is
selected, it can be configured either for Broadcast UDP or Directed UDP.
In case of Broadcast UDP, the iSE will transmit the data to every node on the network.
This can be accomplished if the Remote IP Address is set to 255.255.255.255. The
Broadcast UDP is a practical solution when one iSE needs to communicate with multiple
nodes over the network.
In the case of directed UDP, the iSE will transmit the data to a specific node on the
network. This can be accomplished if the Remote IP Address is set to the IP address of
that specific node.
Server Type
: Continuous mode sends the temperature to the Ethernet, every two
seconds. It is mainly used to send readings to a remote display.
Command mode (default) needs a command to query the iSE to send the reponse back
to querying device.
Forward CR
: Forward CR can be enabled and disabled.
If enabled the iSE will forward a <CR> along with the data to a network host.
Number of Connections
: The range is from 0 to 5. If 0 is selected, the Terminal Server
feature is disabled. This means that no network connection can be made to the sensor
connected to the iSE.
If 1 is selected, only one network connection can be made to the iSE’s.
Any number more than 1 would allow that number of network hosts to read from the iSE
simultaneously (default is 5).
Port
: Port 2000 is the default TCP port for the iSE’s port to which the sensors are connected.
Ports 1000 (used for HTTPget, refer to Section 4.5), 2002, 2003, and 2004 are reserved
for internal use.
Terminal Server usually describes a device that exchanges data between
Ethernet/TCPIP networks and RS-232/RS-485 systems. With this iSE, the data
is obtained digitally from the sensor (irrelevant to RS-232 or RS-485 interface)
and can be accessed from anywhere on the network.
A computer program, such as Mail Notifier, OPC Server, iLog, iFlash or HTTPget
can send TCP requests and obtain readings using the Terminal Server feature.