4.3.5 Configuration (continued)
C) Flash Card Size
Reports the Flash Card size and if it is already recorded, how many locations are filled.
0-Megabytes: Flash Card is not installed or is a bad Flash Card
2-Megabytes, 4-Megabytes, 8-Megabytes: available sizes of Flash Cards.
D) Flash Card/Power Messages Describes the status of the Flash Card and power.
Flash Card Module Malfunctional Flash Card Standby
Flash Card Recording Stopped Flash Card Pre-Recorded
Flash Card Recording Complete Flash Recording On
Flash Card Module Not Initialized Flash Card Corrupt Initialization
Recording Initialized Entries Error
Low Battery Archive Busy
If the battery is not connected or the voltage is low, the statement “Check Battery” will
display. You may also see a message “Archive Busy” if another client is busy
retrieving archive data. Clicking the “Update” button will clear “Archive Busy”.
E) Alarm Setup
There are two alarms which can be assigned to Temperature 1 and Temperature 2
(one alarm to each variable). Selection of Input 1 or Input 2 and Setpoint High
(SP_H) or Setpoint Low (SP_L).
Temperature 1 and Temperature 2: Disable, Low, High, High/Low
Disable: Alarm is disabled (no alarm function)
Low: Alarm is enabled only with Low setpoint
High: Alarm is enabled only with High setpoint
High/Low: Alarm is enabled only with High and Low setpoints
Example: If High/Low is selected, the temperature will be monitored against these two
High and Low values. If temperature goes above High value (082.5), Relay 1 will be
energized, and if temperature goes below Low value (032.5) Relay 1 will be energized.
Input 1 Status: Unlatch, Latch, Clear Latch, Chart SP
Input 2 Status: Unlatch, Latch, Clear Latch, Chart SP
Unlatch: The relay in the iSE will be turned on (closed) when the Temperature 1
and/or Temperature 2 values fall outside of Setpoint range. When the Temperature
1 and/or Temperature 2 values are back within the range, the relay will be turned
off (opened).
Latch: The relay in the iSE will be turned on (closed) when the Temperature 1
and/or Temperature 2 values fall outside of Setpoint range and will remain on until
the user turns it off.
Clear Latch: This option will clear the latch, meaning the activated alarm will be disabled.
Chart SP: Selecting this option will add the setpoints to the ”Chart”.
SP1_L, SP1_H and SP2_L, SP2_H: Editable box for Setpoint values. The format is
nnn.n, it must be followed exactly otherwise the modification will not take place for
any modified item in the Alarm Setup.
All the Alarm Settings will be lost if the unit is power cycled.
In order to keep the Alarm Settings permanently you will need to Start the
Recording. You can then Stop the Recording, if data recording is not needed.
This will ensure the Alarm Settings will be saved on the Flash Memory Card.