Removal of Adhesive Accelerometers 
Many accelerometers and adhesive adapters have been damaged by improper removal. A safe removal 
method is to torque the accelerometer or its adapter with a wrench using the flats provided. Adhesives are 
generally weakest in the shear mode and will yield under steady torque. 
Accelerometer Mounting 
CAUTION 
Never strike an accelerometer to remove it. The trauma will likely damage the 
accelerometer and affect calibration. 
Magnetic Mounting Adapters 
Magnetic mounting adapters can attach accelerometers to ferromagnetic surfaces such as machinery and 
structures where the sensor is to be moved quickly from place to place. The accelerometer is attached to 
the magnet (usually by stud mount) and the assembly is then applied to the surface. 
In general, use magnetic adapters with caution and never beyond 1 kHz. Expect response degradation in 
proportion to accelerometer mass. To ensure the best accuracy from the magnetic mount: 
1.  If possible, attach the magnet to bare metal. Paint will lessen the holding force of the magnet and can 
affect response. 
2.  Clean mounting area to remove oil, grease and foreign matter, which could reduce the contact for a 
strong magnetic bond. 
3.  Select a flat area to achieve maximum surface contact. 
4.  Attach the magnet to the test surface CAREFULLY. Remember that the pull of the magnet rises 
sharply just before contact with the ferromagnetic surface. This force could pull the assembly from 
your hand resulting in a severe metal-to-metal shock. This could over-range the accelerometer beyond 
its maximum shock limit and damage it. 
5.  If possible, calibrate the magnetic/accelerometer assembly. 
Electrical Grounding 
Case-Grounded Design 
In case-grounded designs, the common lead on the impedance-matching circuit is tied to the accelerometer 
case. The accelerometer base/stud assembly forms the signal common and electrically connects to the shell 
of the output connector. Case-grounded accelerometers are connected electrically to any conductive 
surface on which they are mounted. When using these units, take care to avoid errors due to ground noise.
Isolated-Base Design 
To prevent ground-noise error, many accelerometers have a base-isolated design.  The outer case/base of 
the accelerometer is isolated electrically off ground by means of an isolation stud insert. The proprietary 
material used to form the isolation provides strength and stiffness to preserve high-frequency performance. 
DBK4, pg. 14 958293  DBK Option Cards and Modules 