- OMEGA Engineering, Inc. Power Supply User's Guide
Page : 10
11.- ADJUSTMENT AND CALIBRATION PROCEDURE
11.1.- RTD MODELS
A decade box or RTD simulator is required.
1.- Check that the instrument is not powered.
2.- Connect the RTD simulator or decade box to the signal input connector (See paragraph 8.3).
3.- Remove the rear cover taking off all the screws to obtain access to the control board.
(See page 7 paragraph 7).
4.- Power up the instrument with the appropriate power supply and wait until 3 minutes before to
start the adjust.
5.- Adjust the RTD simulator to generate the low signal indicated in TABLE-A.
6.- Display must shows the DISPLAY value indicated in TABLE-A, if not turn the trimmer PT1 located
on the Signal Conditioner (See paragraph 10.1) until display shows the DISPLAY value.
7.- Adjust the RTD simulator to generate the high signal, see TABLE-B.
8.- Display must shows the DISPLAY value indicated in TABLE-B, if not turn the trimmer PT2 located
on the Signal Conditioner (See paragraph 10.1) until display shows the DISPLAY value.
9.- Repeat steps 5 to 8 until the two values are correct.
10.- Check that the linearity remains correct, testing some other points of the range.
11.2.- THERMOCOUPLE MODELS
A Thermocouple simulator with Cold Junction compensation is required.
1.- Check that the instrument is not powered.
2.- Connect the Thermocouple simulator to the signal input connector (See paragraph 8.4).
3.- Remove the rear cover taking off all the screws to obtain access to the control board.
(See page 7 paragraph 7).
4.- Power up the instrument with the appropriate power supply and wait until 3 minutes before to
start the adjust.
5.- Adjust the Thermocouple simulator to generate the low signal indicated in TABLE-A.
6.- Display must shows the DISPLAY value indicated in TABLE-A, if not turn the trimmer PT1located
on the Signal Conditioner (See paragraph 10.2) until display shows te correct value.
7.- Adjust the Thermocouple simulator to generate the high signal indicated in TABLE-B.
8.- Display must shows the DISPLAY value, if not turn the trimmer PT2 located on the Signal
Conditioner (See paragraph 10.2) until obtain the DISPLAY value.
9.- Repeat steps 5 to 8 until the low and the high signals are correct.
10.- Check that the linearity remains correct, testing some other points of the range.
  TABLE - A  TABLE - B
MODEL  LOW SIGNAL DISPLAY MODEL  HIGH SIGNAL DISPLAY
  M1 0°C  000.0    M1  600°C  600.0
 M2 0°F    0000    M2  500°F    0500
 M3 0°F  000.0   M3  200°F  200.0
  J1 0°C    0000   J1  600°C   0600
 J2  32°F    0032  J2    1000°F    1000
  K1 0°C  0000     K1  1200°C  1200
  K2  32°F    0032  K2     2000°F   2000
 T1 0°C   0000   T1  400°C   0400
 T2 0°F    0000   T2  400°F    0400
 E1 0°C   0000  E1  600°C   0600
  E2  32°F    0032  E2     1200°F   1200
  R1  300°C   0300  R1     1700°C   1700
  R2  32°F    0032  R2    3100°F    3100
 S1  200°C  0200  S1    1700°C    1700
  S2  32°F    0032  S2     3100°F   3100
  L1  0°C   0000  L1   600°C   0600










