User manual
Appendix A API Commands
ChartScan User’s Manual A-7
Command
Syntax
Description
D#
Set Relay Make Time
D#make
Specify how long to wait after closing a channel relay before
reading valid data, where 
make
 is the number of 520.833
microsecond intervals.
E
End Calibration Mode
E
Terminate Calibration Mode
E?
Error Status Query
E?
Returns present error condition of the unit with one of the following
error codes:
E000
 - No error has occurred.
E001
 - Invalid device dependent command (IDDC).
E002
 - Invalid device dependent command option (IDDCO).
E004
 - Channel configuration error.
E008
 - Calibration error.
E016
 - Trigger overrun.
E032
 - Open TC or Range Error.
E128
 - Command conflict error.
F
Set Data Format
Fengr,
format
Configure the input and output formats.
engr
 converts the raw data to the engineering units that you
prefer to work with. Valid options are: 
0
 - degrees Celsius
(default), 
1
 - degrees Fahrenheit, 
2
 - degrees Rankine, 
3
 -
degrees Kelvin, and 
4
 - Volts.
format
 places the readings in a format that you prefer to work
with. Valid options are: 
0
 - Engineering Units (default), 
1
 -
Binary (Low Byte/High Byte), 
2
 - Binary (High Byte/Low Byte),
and 
3
 - Counts (ASCII).
F?
Query the data format.
F#
Set Burst Mode
Frequency
F#freq
Select burst mode sampling frequency where 
freq
 is a real
number and where 
38.5
 < 
freq
 < 
20000.0
 Hertz (default).
G
Calibrate Channel Gain
Gchan,
type
Calibrate the correction gain constants for given option card where
chan
 is the channel number such that
1
 < 
chan
 < 
max
,
where max = 128 for ChartScan/1400.
The chassis is 
chan = 0
.
type
 is the channel type.
H
Calibrate Channel
Offset
Hchan
Calibrate the correction offset constants for given option card
where 
chan
 is the channel number such that
1
 < 
chan
 < 
max
, where max = 128 for ChartScan/1400.
The chassis is 
chan = 0
.
I
Set Scan Interval
Inorm, acq
Set the two configurable scan intervals of the unit where 
norm
and 
acq
 are the normal and acquisition scan intervals which
have the form: 
hh:mm:ss.t
 where 
hh
 is hours, 
mm
 is
minutes, 
ss
 is seconds, and t is tenths of a second.
I?
Query scan interval selection.
I#
Digital Input Stamping
I#state
Specify whether or not digital input stamping will be enabled.
state
 may have either of the following values: The default of
0
, which disables digital input stamping, or 
1
 to enable digital
stamping.
J
Calibrate Cold Junction
Offset
Jchan,
type,
temp
Calibrate the cold-junction offsets for cold-junction compensation
of the thermocouple signals where 
chan
 is the channel number
such that 
1
 < 
chan
 < 
max
,
where max = 128 for ChartScan/1400.
type
 is the thermocouple channel type.
temp
 is the reference temperature currently being applied to the
selected channel, with the form 
nnn.n
 in degrees º C.
K
Enter Calibration Mode
Kkey
Enter Calibration Mode if the keyword (or password) is correct,
where 
key
 is the keyword in the form of a 5-digit number
nnnnn
. Therefore the password can be no longer than 5
characters.
K?
Query 
Kkey
 as defined above.