DisplayHi: Defines the high scale for a pen on the display.
StatusBar: For users convenience when viewing the status of Totalizer, Counter, DI or DO.,
the user may enable these items in the StatusBar. For example, after enable Totalizer 13, an
extra bar of Totalizer 1, 2 & 3 will appear at the lower part of display.
Figure 4 – 4
Note:
To illustrate the difference between DisplayHi, DisplayLo and Scale Hi, Scale Low. Here
is a typical example, with input 0-10V, Scale Low=0.00, Scale Hi=100.00, to have better
resolution and vision on Bar, set DisplayLo=0.00, DisplayHi=50.00 so that the Bar
displays from value 0.00 to 50.00.
The decimal point is defined by Scale Hi and Scale Low, and not by DisplayHi, or
DisplayLo.