- Omega Sensor User's Guide
pg. 27 of 29 
Symptom  Possible Cause  Method of Correction 
Flow indication is not 
linear and output too 
low at lower flows 
Attempting to measure 
flows below the specified 
minimum for the unit 
Moisture or liquid in sensor 
Mechanical sensor assembly 
damaged 
Use higher flow rates 
Remove moisture and allow 
sensor to dry out 
Unit must be returned for 
repair (see Section J) 
Flow reading accuracy 
is outside the 
specification 
Gas has a specific gravity 
other than Air at 20°C 
Moisture in sensor 
Mechanical sensor assembly 
damaged 
Pressure drop issue 
Calculate the flow reading 
error / offset as explained in 
section C6 
Allow sensor to dry out. 
Ensure gas is dry. 
Unit must be returned for 
repair (see Section J) 
Correct for the pressure 
drop ( ΔP ) 
Flow indication is fixed 
at one value 
The maximum rated flow for 
the unit is being exceeded. 
Output circuits may be 
damaged 
Reduce the flow 
Unit must be returned for 
repair (see Section J) 