- Omega Sensor User's Guide
pg. 26 of 29 
H.
Troubleshooting Guide 
Symptom  Possible Cause  Method of Correction 
Unit Leaks  Fittings not tight enough. 
Sensor assembly cracked 
Tighten fittings (see section 
B4) 
Unit must be returned for 
repair (see Section J) 
No output signal or 
flow indication 
No power or low power 
No flow passing through 
unit 
Output shorted or 
overloaded 
Improper connection 
Moisture on internal 
electronics 
Output circuitry damaged 
Apply correct power 
Display or output should 
read zero. 
Check electrical connections 
(see section B5) 
Check electrical connections 
(see section B5) and ensure 
connector is correctly 
plugged in. 
Allow unit to dry out in a 
dry environment. 
Unit must be returned for 
repair (see Section J) 
Flow indication 
“bounces” or 
fluctuates 
Flow is actually fluctuating 
Using a pulsatile or piston 
pump that causes a 
pulsating flow of gas 
Not using the fittings 
supplied by the 
manufacturer 
Moisture in sensor 
Turbulence in tubing or 
sensor 
Fluctuating input power 
Mechanical sensor assembly 
damaged 
Correct flow source to 
obtain steady flow 
Switch to a stable source of 
gas 
Use the supplied fittings 
Remove moisture and allow 
unit to dry out 
Straighten the inlet tubing 
or lengthen it to at least 6” 
(150mm) 
Correct the power source or 
change to a regulated 
supply 
Unit must be returned for 
repair (see Section J) 










