- Omega Sensor User's Guide
pg. 17 of 29 
E.
Specifications For Gas Units 
 FLR1000 FLR1000-D FLR1000-BR-D 
Display None 
3.5 Digit LCD, 
0.39” (10mm) high digits 
Accuracy 
(including linearity) 
±3.0% Full Scale* 
Repeatability  ±0.5% Full Scale* 
Pressure Rating  40 psig 
(2.8 bar) 
Temperature Rating  Operating Range: 5 to 55ºC 
Storage Range: 0 to 70ºC 
Temperature 
Sensitivity 
±0.2% F.S.* or less per ºC 
Wetted Materials  Ryton
®
316 Stainless* 
Epoxy 
Glass 
Sapphire 
Brass 
Ryton
®
316 Stainless* 
Epoxy 
Glass 
Sapphire 
O-Ring Material  Standard - Viton
® 
EPDM Optional 
Fitting Material 
(Standard) 
Acetal Brass 
Fitting Material 
Optional 
Acetal, brass, or stainless steel 
Recommended 
Filtration 
25 microns or less 
Compatible Gases  Clean, dry 
0-5VDC Output Signal  Standard 
Minimum 2.5 KΩ load 
Typical Power 
Consumption 
Standard: 12 VDC @ 35 mA (11.5-15 VDC) 
Model 110 Power Adapter 115/230VAC 
Response Time  Typically <30 seconds for 67% of final value 
Reliability 100,000 hours MTBF 
Certifications CE Approved 
89/336/EEC (EN 55011 & EN 50082-1) 
73/23/EEC Low Voltage Directive 
Ratings  IP10 (NEMA 1) 
Warranty  1 year limited 
*Specifications from 20-100% of rated flow. Linearity is best fit straight line. All calibrations 
performed with AIR unless otherwise stated on calibration certificate 