User's Manual
15
5. Wearing safety face protection, turn the
installation tool shaft counterclockwise to withdraw
sensor until the sensor flange contacts the upper
hex nuts.
6. Raise one lower hex and jam nut to bottom of
sensor flange.
7. Close valve, remove bearing plate and tool.
To remove the sensor
8. Wearing safety face protection, cover the
bleed valve with suitable protection (rag, towel,
etc.) and open the bleed valve (ccw rotation) to
relieve internal pressure. Pull sensor up until bleed
valve purges some fluid (indicating sensor is past
1st o-ring seal inside brass sensor nut) then remove
sensor from brass sensor nut/threaded rod
assembly.
Caution: In case of a leaky isolation valve, the
sensor will be under a slight amount of pressure.
Care should be taken when removing the sensor.
Use the bleed valve to relieve this pressure taking
care not to spray fluid on yourself or others.
swivel mount
w/cable port
installation tool
threaded shaft
sensor flange
sensor cable
cap nuts
upper hex nuts
1 lower hex nut
and jam nut
sensor body
installation tool
bearing plate
protector plate
hex nuts
When reinstalling the
sensor:
leave one lower
hex nut in position to guide
sensor to proper height before
opening valve. Return to "H"
dimension height after valve is
opened.