6  Darkfield Condenser Installation and 
Operation Instructions 
1)  Mounting the dry darkfield condenser 
  Rotate the nosepiece and set the 4x objective in 
position. 
  Loosen and raise the focusing rack stop screw to 
the highest position. 
  Turn the coarse focus knob to raise the 
mechanical stage to the highest position without 
contacting the 4X objective. 
  Turn the condenser control knob to raise the 
condenser to the highest position. 
  Loosen the condenser lock thumb screw, pull down the condenser and take it off, 
insert the dry darkfield condenser in, and tighten the condenser lock thumb 
screw (See Fig. 7). 
  Re-adjust the condenser height to its normal position. 
  Re-adjust the mechanical stage height to its normal position. 
  Lower the focusing rack stop screw back and use the bolt to lock it when 
needed. 
Note:   
When raising the mechanical stage, do not make contact with the objective. 
2)  Centering the dry darkfield condenser 
  Turn the 4X objective to the light path. 
  Turn the condenser focus knob to lower the 
condenser till a dark spot showed in the 
viewing field as shown in Fig. 8 (a). 
  Turn the condenser translational centering 
screws to move the dark spot to the center 
as shown in Fig. 8 (b). 
3)  Place the slide on the stage. 
4)  Raise the condenser all the way to the top and then lower it a little bit. 
5)  Follow the procedures in this manual to bring the sample on slide in focus and 
observe. 
6)  Move the condenser up or down slightly to get the best darkfield viewing. 
Note: 
  The dry darkfield condenser is used with the dry objectives only. 
  The dry darkfield condenser works with the 4X, 10X, 40X objectives. 
  The dry darkfield condenser does not work with the 100X oil immersion 
objective. 
(a)                                (b) 
Fig. 7 
Fig. 8 
Condenser Lock   
Thumb Screw 
Centering Screws 
Condenser Holder 