INSTRUCTIONS IX2-ARCEVA ARC ILLUMINATION TIRF SET This instruction manual is for the Olympus Arc Illumination TIRF Set. To ensure the safety, obtain optimum performance and to familiarize yourself fully with the use of this system, we recommend that you study this manual thoroughly before operating the system. As this system employs an arc light source (mercury burner or xenon burner), take special care for safety and please also read the instruction manual for the light source.
Printed on 100% recycled paper with soy ink.
IX2-ARCEVA CONTENTS Correct assembly and adjustments are critical for the system to manifest its full performance. If you are going to assemble the system by yourself, please read chapter 3, “ASSEMBLY” (pages. 8 to 11). For the assemblies of the modules for which instruction manuals are available, refer to their instruction manuals. IMPORTANT — Be sure to read this section for safe use of the equipment. — 1 NOMENCLATURE 1-4 5 2 CONTROLS 6-7 3 ASSEMBLY 8-11 4 OPERATION 12-14 4-1 4-2 Preparation .
IMPORTANT This system has been designed to enable total reflection fluorescence observation using an arc light source (mercury burner or xenon burner). Total reflection illumination is implemented by limiting the illumination projection area to the area corresponding to the objective pupil NA of 1.39 (because the reflectance of cytoplasm is 1.38). }This system offers the following advantages compared to the laser-based total reflection illumination system.
IX2-ARCEVA 1 Getting Ready 1. This manual pertains only to the handling of the total reflection fluorescence illumination system. Please also read the instruction manuals for the IX2 series (IX81/IX71/IX51) microscope and associated options to understand the comprehensive operating instructions of the microscope system. 2. The total reflection fluorescence illumination system is composed of precision instruments. Handle it with care and avoid subjecting it to sudden or severe impact. 3.
13. The aperture iris diaphragm for use in ordinary fluorescence observation is not provided. If it is required to use the aperture iris diaphragm, replace the slit iris diaphragm unit of this system with the aperture iris diaphragm unit of the IX2-RFAL. 14.
IX2-ARCEVA 2 Maintenance and Storage 1. Clean all glass components by wiping gently with gauze. To remove fingerprints or oil smudges, wipe with gauze slightly moistened with a mixture of ether (70%) and alcohol (30%). Since solvents such as ether and alcohol are highly flammable, they must be handled carefully. Be sure to keep these chemicals away from open flames or potential sources of electrical sparks –– for example, electrical equipment that is being switched on or off.
NOMENCLATURE }The modules enclosed in belong to the Arc Illumination TIRF Set. · The illumination field in fluorescence observation is limited to the area of FN 20. · The illumination field in total reflection fluorescence observation is limited in the area of FN 11. L-shaped reflected light illuminator IX2-RFAL Recommended stage IX-SVL-2 Features high operability of iris diaphragm unit.
IX2-ARCEVA CONTROLS Field iris diaphragm centering screws Arc Illumination TIRF Set IX2-ARCEVA Slit focus adjustment screw Adjustment range: ± 1.5 mm Slit iris diaphragm unit Field iris diaphragm lever Slit centering screws Adjustment range: ± 1.5 mm L-shaped reflected light illuminator IX2-RFAL (Note) The two filter sliders and aperture iris diaphragm unit are not used with this system.
4-position excitation filter slider ND filter insertion slots Filter diameter: 32 mm. Thickness: 1 mm or less. Sticker attaching positions x6 Push ring driver Light source side Slider knob Light leak prevention plate Handwritten indication stickers Vacant position Objective side Excitation filter insertion slots Filter diameter: 25 mm. Thickness: 6 mm or less. Wedge prism slider ND filter insertion slot Filter diameter: 32 mm. Thickness: 1 mm or less.
IX2-ARCEVA ASSEMBLY The filters and sliders are very hot during and immediately after use. Confirm that they are cooled down enough before replacing them. # When a commercially available, non-Olympus low reflection ND filter is used, the filter may sometimes be cracked or burnt. 1 Attaching the Click Additional Parts (Fig. 1) 1.
2 ³ @ ² Mounting the Excitation Filter (Figs. 2 to 4) }When the IX2-RFACEVA mirror unit cassette is used, the operation in step 1 below is not necessary because no excitation filter is provided with it. 1. Using the push ring driver ², turn the excitation filter push ring @ of the mirror unit to be mounted in the fluorescence mirror unit counterclockwise and remove the excitation filter ³. 2.
IX2-ARCEVA 4. Write the name (or abbreviation) of the mounted filter on one of the provided handwritten indication stickers and attach it to one of the sticker attaching positions … (position on the left or right of the position in which the filter is mounted). In the above case, attach the sticker to the position on the left of the filter position because the left side comes near the slider inlet when the slider is set on the illuminator. … ³ Fig.
4 Mounting the IX2-RFAL 5 Mounting the UV Cut Plate }Refer to the instruction manual for the IX2 reflected fluorescence system. 6 Mounting the ND Filter }The ND filter attenuates the excitation light and helps delaying color fading of the specimen. Insert the ND filter gently into the ND filter insertion slot of the slider.
IX2-ARCEVA OPERATION 4-1 Preparation 1 (Fig. 5) }The following settings are the original settings required for this system. The details may be variable according to the instruction manual for each lamp housing. @ Field iris diaphragm: Pulled-out position (Closed) ² Slit: Pulled-out position (Slit OUT) ³ Wedge prism: Nearest position (Vacant position) | Excitation filter: B or IB excitation filter engaged. Also the B or IB mirror unit should be engaged in the light path.
ƒ ² | ³ Fig. 7 Fig. 8 13 4. Push in the slit selector lever ² (IN position) and set the wedge prism slider ³ to the intermediate click position to engage the slit and wedge prism in the light path. (Fig. 7) 5. Using the Allen wrench (width across flats 3 mm), turn the two centering screws | alternately so that a crescent slit is visible on the target. Adjust so that the slit is located as close as possible to the perimeter of the target. 6.
IX2-ARCEVA 4-2 Total Reflection Fluorescence Observation · The only objective that can be used in total reflection fluorescence observation is the PlanApoN60XOTIRFM or PlanApo60XOTIRFM. · Since the illumination intensity is lower than with the laser light source, this system is not applicable to molecular fluorescence observation. Its applications are limited to high-contrast observations of tissues near cell membranes and physiochemical phenomena. 1.
TROUBLESHOOTING GUIDE Under certain conditions, performance of the system may be adversely affected by factors other than defects. If problems occur, please review the following list and take remedial action as needed. If you cannot solve the problem after checking the entire list, contact Olympus. Phenomenon a. Total reflection fluorescence cannot be obtained. Cause The PlanApoN60XOTIRFM or PlanApo60XOTIRFM objective is not engaged in the light path. Treatment Page Engage them in the light path.
Shinjuku Monolith, 3-1, Nishi Shinjuku 2-chome, Shinjuku-ku, Tokyo, Japan EC REP Wendenstraße 14-18, 20097 Hamburg, Germany 3500 Corporate Parkway, P.O. Box 610, Center Valley, PA 18034-0610, U.S.A. One Corporate Drive, Orangeburg, NY 10962, U.S.A. 491B River Valley Road, #12-01/04 Valley Point Office Tower, Singapore 248373 31 Gilby Road, Mount Waverley, VIC., 3149, Australia 5301 Blue Lagoon Drive, Suite 290 Miami, FL 33126, U.S.A.