18 VS200 device configuration
7. Select the e.g. 20x phase contrast objective and select the [Use with this
objective] check box.
Refer to the following table for the correct assignment of the different
phase contrast inserts to the correct magnification:
Objective magnification BX3 condenser PH insert
10x U-PH1-S
20x U-PH1-S
40x U-PH2-S
60x U-PH3-S
100x U-PH3-S
8. Select the [BX3 Top Lens] entry from the [Available components] and
select the [Top Lens] > [In] option for the 20x PH objective.
Refer to chapter VS200 LED lamp voltages auf Seite 97 for the correct lamp voltages
for phase contrast acquisition.
- 93 -
Instructions for useOLYMPUS VS200 System