18 VS200 device configuration
18.6.1 ORCA camera adjustments
1. Use the [Acquire] > [Devices] > [Device Settings] command to open the
[Device Settings] dialog box.
2. Select the [BFMono] observation method (OM).
3. Unlock the [BFMono] observation method by clicking on the lock icon.
4. In the [Status] picklist, set the Orca camera to the status [Use].
5. In the [Image type] picklist, change the [Current] entry to the [16-bit gray-
scale] entry.
6. Do this for all other mono camera-related observation methods.