20 Calibrate VS200 using the Olympus Calibration Slide
20.1 Stage limits - Z Axis
The X and Y axes do not have to be calibrated again. However, the Z axis
calibration is dependent on the glass slide thickness. Therefore it is neces-
sary to calibrate the Z axis.
1. In the [Calibrations] dialog box select the [Stage Limits] entry and click the
[Calibrate] button.
2. Proceed with [Next] and follow the instructions of the wizard.
- 103 -
Instructions for useOLYMPUS VS200 System