Use Instructions
Table Of Contents
- 20 Calibrate VS200 using the Olympus Calibration Slide
- 21 Additional calibrations for a fluorescence system
24 Troubleshooting
24 Troubleshooting
24.1 Hardware not available
If the [IX3 Nosepiece:Hardware is not available] error message appears, do the fol-
lowing:
1. Close the VS200 ASW software.
2. Switch off the VS200 system!
3. Check cabling of IX3-RFACA.
24.2 "No camera" error
If you see the following error messages after switching to the monochrome camera
you need to check whether the correct camera is assigned in the observation meth-
ods which use a monochrome camera. See Manual device configuration auf Seite
89.
In case you use an ORCA camera make sure that it is switched on.
24.3 Tray not active
If you see the error message below you possibly inserted a new tray type without hav-
ing it set up correctly in the VS200 ASW software first.