Achten sie darauf, dass Sie während der Reinigung der Immer-
sionsobjektive keine Softwarefunktionen ausführen.
1. To clean lenses, remove dust by blowing them with a commercially avail-
able blower and wipe them lightly with cleaning paper (or a piece of
repeatedly washed gauze). Only if they are stained by fingerprints or oils
should you wipe them using cleaning paper slightly moistened with dehyd-
rated alcohol sold at store
CAUTION
Dehydrated alcohol is highly flammable. Do not expose it to heat or
flame, and do not turn off or on the power switch of various elec-
trical apparatuses since ignition can be induced by just switching
switches on and off. In addition, make sure that the room is well
ventilated.
2. Do not use an organic solution to clean parts other than lenses. If a part is
heavily stained, wipe it with a soft cloth slightly moistened with a diluted
neutral detergent.
Avoid cleaning lenses with the camera or the lenses removed. Doing so
will cause the system to be uncalibrated. All calibrations need to be re-
done.
3. Click the [Clean Objective] button on the start page of the VS200 ASW soft-
ware and follow the instructions in the dialog box.
The objective revolver will move up to grant more space.
It is not necessary to unmount an objective to clean it.