6. Clean the following components with a lint free cleaning cloth.
Left side of the tray hotel
Right side of the tray hotel
The bottom panel of the tray hotel
7. For more pronounced marks, you can lightly wet the cleaning cloth with
water or a mild solvent-free disinfectant or cleaning agent.
8. Push the tray hotel as far back as it will go to the back panel.
9. Manually move the SCARA robot arm to the left.
10. Clean the gripper only with a lint free cleaning cloth.
11. If the loader or individual components were cleaned with a damp cloth,
allow them to dry.
22.3 Cleaning the trays
1. Clean the trays with a lint free cleaning cloth. For more pronounced marks,
you can lightly wet the cleaning cloth with water or a mild solvent-free dis-
infectant or cleaning agent.
22.4 Cleaning the X-Cite Turbo
The X-Cite Turbo light source is not available in all countries.
Cleaning of optics should only be attempted by qualified personnel using appro-
priate fluids and lens paper.
1. To clean the exterior of the unit, use a slightly dampened cloth and a
simple water/ detergent solution only.
22.5 Cleaning the immersion objective
ATTENTION
Damage to the objectives and hardware
Objectives and hardware can get sticky after an immersion medium has been
used.This can damage them.
Clean the immersion objective after each use.
CAUTION
Quetschgefahr durch bewegte Komponenten im Inneren des VS200 Systems
Im Inneren des VS200 Systems bewegen sich mechanische Komponenten.
Wenn Sie bei eingeschaltetem System im Inneren des Systems hantieren,
können Hände und Finger gequetscht und Haare und Kleidung eingezogen
werden.