Use Instructions
Table Of Contents
- 20 Calibrate VS200 using the Olympus Calibration Slide
- 21 Additional calibrations for a fluorescence system
20.4 XY Objective Shift / Parfocality
To carry out a proper XY shift/parfocality calibration use the position of the cal-
ibration slide shown in the screenshot below.
1. In the [Calibrations] dialog box select the [XY Objective Shift / Parfocality]
entry.
2. Click the [Calibrate] button to start the wizard.
- 112 -
Instructions for use
20 Calibrate VS200 using the Olympus Calibration Slide
OLYMPUS VS200 System