- 20 Calibrate VS200 using the Olympus Calibration Slide
- 21 Additional calibrations for a fluorescence system
21 Additional calibrations for a fluorescence system
21.10 Channel-XY-Shift
The channel-XY-shift calibration is only needed if multiple fluorescence fil-
ters or dichroic mirrors in the IX3-RFACA are used.
As this calibration depends on the customer's specific system configuration
and customer's samples, the Olympus employee installing the system
should ask for a dedicated sample slide. If such a sample slide is not avail-
able, the customer itself can do the calibration at any time.
It is important to correct a XY pixel shift induced by the different filter cubes.
1. In the [Calibrations] dialog box select the [Channel-XY-Shift] entry.
2. Click the [Calibrate] button.