Use Instructions
Table Of Contents
- 20 Calibrate VS200 using the Olympus Calibration Slide
- 21 Additional calibrations for a fluorescence system
The system will activate the live mode with the 4x objective used.
5. Use the [Stage Navigator] tool window to navigate to the 20x check-
erboard and focus the sample.
6. Defocus by going down (-100 to 200µm in Z) until you see a homogenous
image (no checkerboards) anymore.
In focus Defocused
- 166 -
Instructions for use
21 Additional calibrations for a fluorescence system
OLYMPUS VS200 System