- 20 Calibrate VS200 using the Olympus Calibration Slide
- 21 Additional calibrations for a fluorescence system
21 Additional calibrations for a fluorescence system
After the acquisition process is finished, the calibration process auto-
matically moves to the next objective.
IMPORTANT
The shading correction for objectives with a magnification equal
to or higher than 10x should be done on a sample glass slide with
cover slip. You will receive good results if you use the VS-cal-
ibration slide.
8. Proceed with all other objectives in the same way.