9. Switch to the [View] layout, go to the [Measurement and ROI] tab and
select either the [Vertical Line] button or the [Horizontal Line] button.
10. Measure the shift between the cross hair and the center structure and note
the X and Y values.
In this case the monochrome camera is shifted up an to the left. To
adjust the shift the cross hair needs to be shifted’ to the right (+X) and
down (+Y).
11. Remember these values (+3.86µm in X and +18.00µm in Y) to enter them
in the [Camera Shift] dialog box in the following steps.
- 148 -