- 20 Calibrate VS200 using the Olympus Calibration Slide
- 21 Additional calibrations for a fluorescence system
21.2 Camera-To-Stage Rotation
The standard for the camera-to-stage rotation is +/- 0.1°. However the
closer you get to zero the better.
1. In the [Calibrations] dialog box select the [Camera-To-Stage Rotation]
entry and click the [Calibrate] button.
2. Use the stage navigator to move to the center of the calibration slide.
- 144 -
Instructions for use
21 Additional calibrations for a fluorescence system
OLYMPUS VS200 System