- 20 Calibrate VS200 using the Olympus Calibration Slide
- 21 Additional calibrations for a fluorescence system
3. Select an area on the calibration slide were it is possible to focus with both
the 40x and 2x objectives like in the screenshot below.
4. Click the [Next] button to continue.
If the focus distance between the objectives differs too much, the wizard
tells you to either shorten or stretch the camera adapter.
5. To adjust the distance, loosen the [LOCK] hex socket screw slightly and use
the [FOCUS] hex socket screw to stretch or shorten the camera adapter. To
turn the [FOCUS] screw use the angled hex key.
If the value is positive, turn slightly clockwise.
If the value is negative, turn slightly counter clockwise.
6. Tighten the [LOCK] screw.
7. Click the [Next] button in the [Camera Adapter Calibration] dialog box.
8. Repeat until the value meets the standard (+/- 20µm or better).