Use Instructions

After the acquisition is complete, the calibration process automatically
moves to the next objective.
The shading correction for objectives with a magnification equal
to or higher than 10x should be done on a sample glass slide with
cover slip. You will receive good results if you use the VS-cal-
ibration slide.
7. Focus on the part of the slide that contains the sample and subsequently
move to a very clean area to acquire the shading image.
8. Proceed with all other objectives in the same way.
9. Click the [Finish] button to finalize the calibration process.
- 134 -
Instructions for use
20 Calibrate VS200 using the Olympus Calibration Slide