Use Instructions

20 Calibrate VS200 using the Olympus Calibration Slide
20.6 Shading correction of the label area
20.6.1 Preparation of calibration slide
For the shading correction calibration of the label scan you need to prepare a
proper slide.
If a label printer is used, take an empty sticker and stick it on a normal glass slide as
shown in the image below. If the customer does not use a label printer take a piece
of white copy paper and glue it onto the slide.
Example of a slide with an empty sticker
1. Insert the VS-calibration slide into position 3 of the slide tray and the pre-
pared label-shading-correction slide in position 2.
- 123 -
Instructions for useOLYMPUS VS200 System