20 Calibrate VS200 using the Olympus Calibration Slide
8. Click the [Next] button to start the image acquisition process for the 2x
objective.
After the acquisition is complete, the calibration process automatically
moves to the next objective.
The shading correction for objectives with a magnification equal
to or higher than 10x should be done on a sample glass slide with
cover slip. You will receive good results if you use the VS-cal-
ibration slide.
9. Proceed with all other objectives in the same way.