5.4 Filter Replacement Procedure:
1. Turn unit off and disconnect the power cord.
2. Unlatch both filter cover catches and remove the filter cover. (see below)
3. Replace the filters. (Pre-filter/main filter). Ensure the main filter is placed
inside the alignment pins so that the interlock switch is depressed.
4. Replace the filter cover and ensure a proper fit to the bottom housing.
Hook both latches into the filter cover and press down carefully to
engage each latch to the base.
Caution! Do not use force. If latches do not close, check proper
seating of filter cover on bottom housing.
5. Reconnect power cord and turn system on.
6. If the unit does not operate, refer to the troubleshooting section.
Notice: Always use new replacement filters!
5.5 Filter Disposal
The gas filter collects some organic materials. Note: care in disposal may be
required. A used filter may also contain toxic materials and inhalation of dust
particles or exposure to skin must be avoided. Please consult the MSDS
Sheets for the material used at your facility and follow the prescribed handling
instructions.
Step 2:
Unhook
Latch
Step 1:
Pull & Lift
Latch
Place Filter
inside the
Alignment Pins