CD Indicators Ohaus Corporation 29 Hanover Road Florham Park, NJ 07932-0900 INSTRUCTION MANUAL Models CD-31 and CD-33 Indicators 1
CD Indicators Ohaus Corporation, 29 Hanover Road, Florham Park, New Jersey, 07932, USA Declaration of Conformity We, Ohaus Corporation, declare under our sole responsibility that the instrument models listed below marked with “CE” - are in conformity with the directives and standards mentioned. Konformitätserkärung Wir, die Ohaus Corporation, erklären in alleiniger Verantwortung, dass die untenstehenden Waagentypen, instrument mit “CE” - mit den genannten Richtlinien und Normen übereinstimmen.
CD Indicators NOTE: THIS EQUIPMENT HAS BEEN TESTED AND FOUND TO COMPLY WITH THE LIMITS FOR A CLASS A DIGITAL DEVICE, PURSUANT TO PART 15 OF THE FCC RULES. THESE LIMITS ARE DESIGNED TO PROVIDE REASONABLE PROTECTION AGAINST HARMFUL INTERFERENCE WHEN THE EQUIPMENT IS OPERATED IN A COMMERCIAL ENVIRONMENT. THIS EQUIPMENT GENERATES, USES, AND CAN RADIATE RADIO FREQUENCY ENERGY AND, IF NOT INSTALLED AND USED IN ACCORDANCE WITH THE INSTRUCTION MANUAL, MAY CAUSE HARMFUL INTERFERENCE TO RADIO COMMUNICATIONS.
CD Indicators TABLE OF CONTENTS OVERVIEW OF CONTROLS ........................................................................................................................ 1 OVERVIEW OF DISPLAY INDICATORS ...................................................................................................... 2 1. GETTING TO KNOW YOUR INDICATOR ............................................................................................................ 3 1.1 Introduction3 2. INSTALLATION ....................
CD Indicators TABLE OF CONTENTS (Cont.) 4. OPERATING YOUR INDICATOR ....................................................................................................................... 23 4.1 Calibration23 4.1.1 Span Calibration ..................................................................................................................................... 24 4.1.2 User Calibration .......................................................................................................................
CD Indicators OVERVIEW OF CONTROLS 1 8 2 9 3 4 5 10 6 11 7 No. Designation Function 1 Mode button Selects standard weighing, percent, parts counting, check weighing and animal weighing. 2 Setup button Selects various submenus: scale, calibration, date, time, readout, GMP data, GMP set, print, RS232, LFT, mode, units, global, custom. 3 G/N/T button Selects Gross, Net or Tare weights. 4 Units button Selects weighing units: grams, kilograms, pounds, custom, parts counting, percent.
CD Indicators OVERVIEW OF DISPLAY INDICATORS 6 7 8 9 11 10 5 4 12 3 13 2 14 1 18 No. Use (Pointer Group) keys to change - used to prompt the user while navigating through the menu system. 2 Standard (7) segment numeric characters. Seven characters are available and are used for displaying weight values. 3 Center of Zero - Indicates display is within ± 1/4d of zero. 4 P - Displayed if preset tare weight. 5 T - Displayed with tare weight.
CD Indicators 1. GETTING TO KNOW YOUR INDICATOR 1.1 Introduction Thank you for deciding to purchase a CD-31/CD-33 Indicator from Ohaus. It offers a high level of operating convenience and useful functions to make accurate measurements. A unique LCD panel has a large 7 digit, 7 segment display which indicates the weight value of an item being measured and a 7 digit British Flag display (14 segments) which spells out items selected in the submenus.
CD Indicators 2.3 Connecting the Indicator to a Scale Base • Turn the Indicator over. NOTE:Check the wiring instructions of your scale base. • The Terminal Block located at the left side of the access area is used to connect a scale base using a 4-wire cable and ground. If a 6-wire cable is supplied with the scale base, tie the +SENSE and +EXC leads together and connect to the +EXC connector. Tie the -SENSE and -EXC leads together and connect it to the -EXC connector on the Indicator.
CD Indicators 2.5 Connecting Power Connect the AC Adapter supplied to the connector located at the bottom of the Indicator. Route the wire from the AC Adapter through the channel in the bottom of the Indicator. This will relieve stress on the connector and hold the wire in place. 2.5.1 Cautionary Notes • Models CD-31 and CD-33 Indicators must not be operated in hazardous areas. • Before connecting the AC adapter, verify that the voltage printed on it corresponds to the local mains voltage.
CD Indicators 2.6 Scale Base Setup In this section, you will enter the menu for the first time. Do not worry if you are unfamiliar with the function of the buttons on the panel, the display provides the necessary coaching as you go along. Before the Indicator can be used, the parameters for the scale base being used with the Indicator must be set first in order to calibrate and perform any functions.
CD Indicators 2.6 Scale Base Setup (Cont.) SETUP SUBMENUS SCALE Zero Calunit Grad 2%, 18%, 100% kg, lb .0005, .001, .002, .005, .01....100 Enter 7 digits ON/OFF Full SC Lock Exit .5d, 1d, 2d, 5d OFF, .
CD Indicators 2.6 Scale Base Setup (Cont.) Review the specifications of the scale base to be used with the Indicator. Make sure the settings you select in the indicator are compatible with the scale base. Alternate displays are shown in this procedure to save space. Procedure • Press the Setup button, SCALE is displayed. ZERO Zero specifies the percentage of full capacity load (2%, 18% or 100%) that may be cleared by pressing >0<. Limits may be required in certain applications.
CD Indicators 2.6 Scale Base Setup (Cont.) • Press Enter button, LOCK OFF is displayed, select either ON or OFF using or buttons, then press Enter button SAVED is momentarilydisplayed followed by EXIT. When LOCK OFF= jumper on the bottom is in place, Setup is unlocked. When LOCK ON= jumper was removed, setup is locked. • Press Enter button, Indicator returns to a weigh mode. The Indicator is now matched up with the scale base and the Indicator parameters may now be set and calibrated.
CD Indicators 3. SETTING UP YOUR INDICATOR 3.1 Setting Date and Time Your Indicator provides date and time data which can be viewed on a computer or printed out on an external printer. When you put your new instrument into operation for the first time, you should enter the current date and the time. These settings are retained as long as the Indicator remains connected to a power source. Date Procedure Date is a feature which enables the Indicator to be set to a U.S.A.
CD Indicators 3.2 Readout The Readout menu is used to adapt the Indicator to environmental conditions. It contains five submenus: Stable, Auto 0, Filter, Sleep, Lock and Exit. Lock enables you to lock the settings. Review the settings available below before proceeding. Stability The stability range specifies the weighing results and must be within a preset tolerance limit for a certain time to turn the stability indicator ON.
CD Indicators 3.2 Readout (Cont.) Procedure • • • • • • • • • • • • • • • • • • • 17 Press the Setup button, SCALE is displayed. button until READOUT is displayed. Press Press Enter button, STABLE is displayed. Press Enter button, STABLE .5d is displayed. Press or button and select either .5d, 1d, 2d, or 5d. 1d is the default setting. Press Enter button, SAVED is momentarily displayed. Display advances to AUTO 0. Press Enter button, AUTO 0 .5d is displayed. or button and select either .OFF, .
CD Indicators 3.3 Good Manufacturing Practices (GMP) Data The GMP Data submenu enables the storage of a user identification number (7 digits), a project number (7 digits), and a Scale ID number (7 digits). When entered into the Indicator, the identification number, project number and Scale ID are available when printing providing they are turned on in the GMP Set submenu. A lock setting is also available which locks the settings.
CD Indicators 3.4 Good Manufacturing Practices (GMP) Set Good Manufacturing Practices (GMP) Set submenu allows the selection of: Time, Scale Identification Number, User Identification Number, Project Number, Difference and Name data to be printed. When the selected items are set to ON, these items are displayed. The default setting is OFF.
CD Indicators 3.5 Print The Print menu provides a number of options which can be turned ON or OFF. It contains 8 submenus: Auto Print, Inter, Stable, Numeric, GMPCont, GMP Tare, Reference and Lock. See descriptions below. Procedure • Press the Setup button, SCALE is displayed. • Press or button until PRINT is displayed. • Press Enter button to continue. • Press or button until either AUTOPRT, INTER, STABLE, NUMERIC, GMPCONT, GMPTARE, REFEREN, LOCK or EXIT is displayed.
CD Indicators 3.6 RS232 The RS232 menu provides communication parameters which can be set to accommodate external printers or computers. It contains four submenus: Baud rate, Parity, Data and Lock. See descriptions below. Procedure • Press the Setup button, SCALE is displayed. • Press or button until RS232 is displayed. • Press Enter button. • Press or button until either BAUD, PARITY, DATA, STOP, LOCK or EXIT is displayed. • Press Enter button to save setting.
CD Indicators 3.7 Legal for Trade (LFT) Legal for Trade (LFT) is a software controlled option which can be set to LFT LOCK. When LFT LOCK is set ON, certain items in the Calibration, Readout, Print, Mode and Units menus are automatically preset and locked. This is done to permit the Indicator to operate in a legal for trade application and works in conjunction with a Lock Switch (jumper). Default setting is UNLOCKED, which limits display resolution to 1:10,000. See default table.
CD Indicators 3.8 Mode The Mode submenu permits the selection of 6 modes which can be turned ON or OFF. These modes are: Weigh, Percent, Count, Animal, Checkweigh and Lock. Weigh is turned ON and all others have a default setting of OFF. When any of the modes are turned ON, they can be selected for operation from the Mode button. Procedure • Press the Setup button, SCALE is displayed. • Press or button until MODE is displayed. • Press Enter button.
CD Indicators 3.9 Units The Units submenu permits the selection of the measuring units which can be turned ON or OFF and locked. Procedure • Press the Setup button, SCALE is displayed. • Press or button until UNITS is displayed. • Press Enter button to continue. • Press or button until desired measuring unit is displayed. • Press Enter button to continue. • Press or button and select either ON or OFF. • Press Enter button to save setting. • Press or button to select next item or EXIT.
CD Indicators 3.10 Global (Cont.) List This submenu can be used to output a listing of current menu settings via the RS232 interface. When YES is selected, all menu settings will be output either to an external printer or computer. To use this feature, your Indicator must be connected to a computer or printer. The default setting is OFF. The partial sample shown, indicates the status in the menus.
CD Indicators 3.11 Custom Unit Custom Unit is enabled when Custom Unit Setup under Units Menu is turned ON. This feature can be used to create your own custom weighing unit. It permits entering a conversion factor which the scale will use to convert kg to the desired unit of measure. Conversion Weight Weight SCIENTIFIC NOTATION Factor x in = in Number kg custom unit Between Conversion factors are expressed in scientific notation and entered Conv. 0.1 and Power Maninto the balance in three parts: Factor 1.
CD Indicators 3.11 Custom Unit (Cont.) Procedure (Cont.) • Press Enter button, SAVED is momementary displayed followed by LSD. There are 6 LSD (least significant settings you can choose from (see table). LSD’s LSD .5 Adds one decimal place display counts by 5’s. LSD 1 Display counts by 1’s. LSD 2 Display counts by 2’s. LSD 5 Display counts by 5’s. LSD 10 Display counts by 10’s. LSD 100 Display counts by 100’s. • Press Enter button, LSD 1 is displayed. • Press or button and select LSD value.
CD Indicators 4. OPERATING YOUR INDICATOR 4.1 Calibration Model CD-31 and CD-33 Indicators offer a choice of four calibration methods: Span, User, Linearity, and CalTestTM. • Span - Span calibration ensures that the Indicator reads correctly within specifications using a weight value of 100% of capacity. • User - User calibration is a method where the Indicator can be calibrated using a mass between 10% and 100% of capacity and by entering that numeric value into the Indicator.
CD Indicators 4.1.1 Span Calibration Span calibration requires a mass equal to 100% of the scale base capacity. Procedure • Press the Setup button, SCALE is displayed. • Press button until CAL is displayed. • Press Enter button, CAL TYPE is displayed. • Press Enter button, CAL TYPE SPAN is displayed. • Press Enter button, CLR PAN is displayed. Clear the pan. • Press Enter button, WORKING is momentarily displayed, then PUT WT is displayed (capacity of base) on the pan.
CD Indicators 4.1.2 User Calibration User calibration is used when it is desired to calibrate the Indicator using a mass of known value from 10% to 100% of the capacity of the scale base. To use this calibration feature, proceed as follows: Procedure • Press the Setup button, SCALE is displayed. • Press button UNTIL CAL is displayed. • Press Enter button, CAL TYPE is displayed. • Press Enter button, SPAN is displayed. • Press button to select user calibration. CAL TYPE USEr is displayed.
CD Indicators 4.1.3 Linearity Calibration Linearity calibration utilizes three calibration points; one at zero, center span and full span. This method minimizes deviation between actual and displayed weights within the Indicator's weighing range. Three weight values are used; zero, a weight value at or near midpoint of the Indicator's weighing range and a weight value at specified capacity. Sample display illustrates a 40 kg scale base. Procedure • Press the Setup button, SCALE is displayed.
CD Indicators 4.1.4 Calibration Test Calibration test feature allows a check of a known calibration mass against the last stored calibration information in the indicator. Sample display illustrates a 40 kg scale base using a 10kg mass. Procedure • Press the Setup button, SCALE is displayed. • Press button UNTIL CAL is displayed. • Press Enter button, CAL TYPE is displayed. • Press Enter button, SPAN is displayed. • Press button to select calibration test,CAL TYPE CALtESt is displayed.
CD Indicators 4.1.5 Calibration GMP Printout If any option in the GMP Set Menu is turned On, GMP automatically prints data after calibration is completed. Span Calibration Printout - - - - - SPAN CAL - - - - - 12/01/99 1:10:00 PM Scale Id 1234 Cal: 40kg Old: 39.99kg Dif: 0.01kg Wt. Ref...................................... USER NO 2056853 PROJ NO 100012 Name........................................
CD Indicators 4.2 Weighing The Indicator is shipped with kilograms only enabled. When the Indicator is to be used with other Type Approved/Legal for Trade units of measure, the desired unit must be enabled. Refer to paragraph 3.9 to enable other measuring units. The Indicator is equipped with a G/N/T button which allows the viewing of gross weight, net weight and tare weight. A Tare button allows taring the weight of container, and a separate >0< zero button, zero's the Indicator display reading.
CD Indicators 4.2.2 Preset Taring Preset taring is used to enter a tared weight via the numerical keypad. If a container's weight is known, the Indicator may be preset using the numerical keypad to enter the weight. When using this feature, a time constraint of 4 seconds is imposed to start entering data, otherwise the Indicator defaults back to a weighing mode. Procedure • Press >0< button with no load on the pan to set the Indicator reading to zero.
CD Indicators 4.3 Percent Weighing Percent Weighing is enabled only when Percent is turned ON in the Mode submenu under Setup. Percent weighing permits you to place a reference load on the pan; then view other loads as a percentage of the reference. The load you place on the pan as a reference may be displayed as any percentage you select from 5% to 100% (in 1% increments). 100% does not necessarily have to represent the reference load.
CD Indicators 4.4 Parts Counting Parts Counting is enabled only when Count is turned ON in the Mode submenu under Setup and selected with the Mode button. In the parts counting mode, the Indicator displays the quantity of parts you place on the pan. Since the Indicator determines the quantity based on the average weight of a single part, all parts must be reasonably uniform in weight. Procedure • Press the Mode button. • Press or button until COUNT is displayed.
CD Indicators 4.5 Animal Weighing Animal Weighing is enabled only when Animal is turned ON in the Mode submenu under Setup. Procedure • Press the Mode button, WEIGH is displayed. • Press or button until ANIMAL is displayed. • Press Enter button to continue, LEVEL is displayed. • Press or button to change animal weighing level, 0, 1, 2 or 3. 0 level represents an inactive subject, 3 is used for a very active subject. • Press Enter button to continue, AUTO is displayed.
CD Indicators 4.6 Checkweigh Checkweigh is enabled only when Checkweigh is turned ON in the Mode submenu under setup. Checkweigh permits entering a minimum weight value and a maximum weight value though the numerical keypad or arrow buttons. You can establish minimum and maximum values for repeitive weighing of identical objects. This is suited for production where a large number of items have to be checked against preset parameters. Procedure • Press the Mode button, WEIGH is displayed.
CD Indicators 4.7 Printing Data Printing data to an external computer or printer requires that the communications parameters in the Setup menu, Print options and communication parameters be set first. Refer to paragraph 3 Setting Up Your Indicator. When time, date, GMP Data, Print and RS232 data is set, the printout data will reflect whatever parameters are turned on. Procedure • Press the Print button. Printing to an external printer or computer will occur each time the Print button is pressed.
CD Indicators 4.7 Printing Data (Cont.) 4.7.2 Output Formats (Cont.) RS232 COMMAND TABLE Command Character ? nnnA C F xl L xM P xSL T N V EscV xZ x% x# EscR # % xF Description Print current mode: g, kg, custm, %, # Set Auto Print feature to "nnnn". nnnn=0 Turns feature OFF nnnn=S Output on stability nnnn=C Output is continuous nnnn=1-3600 Sets Auto Print interval Begin span calibration. Print current function. Set Averaging Filter to "x", where x = 0 to 3. If LFT, level 0 to 1.
CD Indicators 4.7 Printing Data (Cont.) 4.7.2 Output Formats (Cont.) RS232 COMMAND TABLE Command Character xAW xE xT ID xID AC xUC LE xS TIME mm/dd/yy SETDATE hh:mm:ss SETTIME DATE SW Z NG EscM E Description 1=percent 2=parts counting 3=animal weighing 4= checkweigh Set Animal Level to "x", where x= 0 to 3. 0= minimum filtering 1= 2= 3= maximum filtering Turn Auto restart in Animal mode on or off. Where x=0 is off and x=1 is on. Set tare weight. Uses real number "x" as current tare weight.
CD Indicators 5 CARE AND MAINTENANCE To keep the Indicator operating properly, it should be kept clean and free from foreign material. If necessary, a cloth dampened with a mild detergent may be used. 5.1 Troubleshooting SYMPTOM PROBABLE CAUSE(S) REMEDY Unit will not turn on. Power cord not plugged in or properly connected to Indicator. Check power cord connections. Incorrect weight reading. Indicator was not re-zeroed before weighing. Press >0< with no weight on the pan, then weigh item.
CD Indicators 5.2 Error Codes List Error Codes List The following list describes the various error codes and which can appear on the display and the suggested remedy. Tare Errors 2.0 2.1 2.2 Tare timeout - Indicator unstable during tare attempt. Preset tare timeout - (more than 4 seconds between key presses). Zero timeout - Indicator unstable during zero attempt. Calibration Errors 3.0 Improper or missing mass used for calibration. Recalibrate with correct masses. User Errors 4.4 5.1 5.2 5.3 7.0 7.1 7.
CD Indicators 5.3 Service Information If the Troubleshooting section does not resolve or describe your problem, you will need to contact an authorized Ohaus Service Agent. For Service assistance in the United States, please call Aftermarket, Ohaus Corporation toll-free at (800) 526-0659. An Ohaus Product Service Specialist will be available to help you. 5.4 Replacement parts Description Ohaus Part No.
CD Indicators LIMITED WARRANTY Ohaus products are warranted against defects in materials and workmanship from the date of delivery through the duration of the warranty period. During the warranty period, Ohaus will repair, or, at its option, replace any component(s) that proves to be defective at no charge, provided that the product is returned, freight prepaid, to Ohaus.
CD Indicators Ohaus Corporation 29 Hanover Road, Florham Park, NJ 07932, USA Tel: (973) 377-9000, Fax: (973) 593-0359 With offices worldwide. P/N 80'250'639 R1099 © Ohaus Corporation 1999, all rights reserved.