Install User guide

016-700000-000-02-201301 43
Appendix C
NIRQuest512-2.5 and
NIRQuest256-2.5 Sensitivity
When configuring a system for operation out to 2.5µm it is important that you consider the following
details:
Short Integration Times: For the detector to be sensitive out to 2.5µm, the detector’s band gap
energy must be small. Unfortunately, this raises the absolute level of the detectors dark signal.
Typical dark signals at 400ms (NIRQuest512-2.5) and 135ms (NIRQuest256-2.5) are shown
below.
Fiber Selection: For maximum signal intensity, alternative fiber materials should be used for
wavelengths greater than 2.2µm.