User Manual
Model
TOS-6/0925
Dimensions
56mm (L) x 30mm (W) x 30mm (D)
Anti-electric Shock Type
Internally Powered Equipment
Anti-electric Shock Degree
Type BF Equipment
EMC
Type B
Mode of Operation
Continuous Operation
Enclosure Degree of ingress protection: IP22
*IP22 means the outer body of the product can withstand the water dropping to the surface when the
body deviate 15 degree from horizontal surface.
Power
2 AAA (Included)
Power Consumption
25- 50mA (Normal)
Environmental
Operating Temperature
About +50~+104°F / +10~ +40°C
Storage Temperature
About .-4~+140°F / -20~+60°C
Relative Humidity
15% - 95% non-condensing
Air Pressure
About 70Kpa to 106Kpa
ALM Limit Default Value
Parameter
Value
Hemoglobin Saturation
Limit: 100 (Upper) 94 (Bottom)
Pulse Rate
Limit: 130 (Upper) 50 (Bottom)
Electronics Parameters
Parameter
Value
Hemoglobin Saturation Display
0-99%
Pulse Rate Display
30-250BPM
Resolution
Hemoglobin
Saturation
1%
Pulse Rate
1 BPM
Measure Accuracy:
Hemoglobin
+ 2% (70% -100%)
Saturation
Unspecified (<70%)
Pulse Rate
+ 3 BPM
PI
Display
0-20%
Resolution
0.1%
Measure Accuracy
0-1%: 0.1%
1-20%: 1%
Probe TFT Specification
Wave Length
Radiant Power
Red
660±2 nm
1.8 mW
Infra RED
905±2 nm
2.0 mW