User Manual

Guidance and manufacturer´s declaration electromagnetic immunity
for EQUIPMENT and SYSTEM that are not LIFE-SUPPORTING
This device is intended for use in the electromagnetic environment specified below. The user of this device
should assure that it is used in such an environment.
Immunity Test
IEC 60601
Test level
Compliance
Level
Electromagnetic Environment - Guidance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to
80 MHz
N/A
Portable and mobile RF communications equipment should
be used no closer to any part of theA310 Fingertip Pulse
Oximeter, including cables, than the recommended
separation distance calculated from the equation
applicable to the frequency of the transmitter.
Radiated RF IEC
61000-4-3
6Vrms in
ISM band
between
150 kHz to
80 MHz
Recommended separation distance
P
V
d ]
5.3
[
1
=
80 MHz to
2.7 GHz
10 V/m
P
E
d ]
5.3
[
1
=
80 MHz to 800 MHz
P
E
d ]
7
[
1
=
800 MHz to 2.5 GHz
where p is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in metres (m).b
Field strengths from fixed RF transmitters, as determined
by an electromagnetic site survey, a should be less than the
compliance level in each frequency range. b
Interference may occur in the vicinity of equipment marked
with the following symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic is affected by absorption and
reflection from structures, objects and people.
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which
the A310 Fingertip Pulse Oximeter is used exceeds the applicable RF compliance level above, the A310
Fingertip Pulse Oximeter should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as reorienting or relocating the A310 Fingertip Pulse
Oximeter.
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m.