User Manual

Radiated RF
IEC 61000-4-3
10V/m
80MHz to
2.7GHz
10V/m
80MHz to
2.7GHz
9V/m,
27V/m,
28V/m
385-5785
MHz
9V/m,
27V/m,
28V/m
385-5785
MHz
Portable RF communications
equipment
(including peripherals such as
antenna cables and external
antennas) should be used no
closer than 30 cm (12 inches)
to any part of this device,
including cables specified by
the manufacturer. Otherwise,
degradation of the
performance of this
equipment could result.
NOTE: UT is the a.c. mains voltage prior to application of the test level.
Immunity
Test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Power
frequency
(50Hz/60Hz)
magnetic field
IEC 61000-4-8
± 8 kV
contact
± 2 kV,
± 4 kV,
± 8 kV,
± 15 kV air
± 8 kV
contact
± 2 kV,
± 4 kV,
± 8 kV,
± 15 kV air
30 A/m 30 A/m
Floors should be wood,
concrete, or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%
Magnetic fields from common
appliances are not expected
to affect the device.
IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment - Guidance