SET UP INSTRUCTIONS FOR MORTUARY REFRIGERATORS / FREEZERS FOR MODELS NMR1 NMR2 NMR3 NMRB1 NMRB2 NMF1 NMF2 NMF3 Please Read instructions prior to using cabinet
Set up instructions for Laboratory Refrigerators Dear customer, Thank you for choosing our Mortuary refrigerator / Freezer, to ensure the cabinet is installed correctly please follow the following steps. Important safety instructions 1. To obtain full use of the cabinet, we recommend reading this instruction manual. 2. It is the user’s responsibility to operate the appliance in accordance with the instructions given. 3. Contact your dealer immediately in case of any malfunctions. 4.
Light switch has been placed either on control panel or next to controller. Door Lock Is fitted on side of door. Remove plastic film on stainless steel, roll out stretchers, clean interior and exterior of cabinet. If necessary with warm water and detergent. Our mortuary refrigerators / freezers can be either with built in motor or remote (motor is placed externally of building or in a plant room). Electrical Connection. 1.
It is essential that condenser be kept clean at all times. Condenser is similar to car radiator. Turn power supply off and brush in a vertical motion to remove any carpet fibres, fluff etc. this can vary from one location to another; it may be necessary to do this every 3 months. Cleaning of refrigerator. It will be necessary to clean the interior and exterior of your cabinet from time to time. Do not use any harsh chemicals or scorers, simply detergent, warm water and a cloth.
Alarm set points Low alarm -1°C High alarm +10°C Delay on alarm is 30 minutes Temperature set point 2 - 5℃ How to Check Hi / Low Readings • Press Hi/low button your max temperature will appear while flashing. • While Hi Temp is flashing, press the Hi/Low button again and your low temperature will display To RESET Hi / Low temperature • Press hi/low button, hold in until beeps approx.