Manual
60
Sound Level Meter
How to Measure the Correction Value
Play a pink noise signal at the typical sound pressure level of •
the live event. (signal source Minirator, NTi Audio Test CD)
Select the correction page •
.
Search the audience area for the location with the highest •
sound pressure level and position the XL2 Analyzer.
Turn the rotary wheel •
to select the parameter RUN next
to Audience and press enter
.
The XL2 measures the sound level at the loudest position
for 5 seconds. The timer counts down to zero.
Wait until the measurement is completed.•
Position the XL2 at the measurement position.•
Turn the rotary wheel •
to select the parameter RUN next
to Measure and press enter
.
The XL2 measures the sound level at the measurement
position for 5 seconds. The timer counts down to zero.