Manual

235
Appendix
L
Aeq
+ k1 Time-Average Sound Level with Correction Value
The actual measurement position often differs to the loudest posi-
tion of the live event. Thus the correction value k1 is the correc-
tion value of L
Aeq
between the loudest position and the actual
measurement position, measured using pink noise. The correc-
tion value k1 is measured or manually set in the page K
SET of the
sound level meter function.
->
->
L
Aeqxx
Moving Time-Average Sound Level
Measurement of the moving sound level over time Leq with A-
weighting frequency and F-weighting time responses; the param-
eter xx is user-defined from five seconds up to one hour. Four
individual levels can be configured, e.g.
5 seconds for live sound monitoring•
10 minutes (in accordance with the French standard)•
15 minutes (in accordance with the British standard)•
60 minutes (in accordance with SLV)•
->
->
->