Manual

228
Appendix
These levels are shown in the numeric result page:
L
Aeq5”
+k1 Time-averaged sound pressure level L
Aeq
with correction value k1 and a moving time
window of 5 seconds.
L
Aeq60’
+k1 Time-averaged sound pressure level L
Aeq
with correction value k1 and a moving time
window of 60 minutes.
L
AFmax
+k1 Maximum sound pressure level L
AFmax
with correction value k1.
The following sound levels are displayed in the real-time ana-
lyzer page:
L
ZFhold3
For tracing of feedback frequencies. The
peak hold time is set to 3 seconds.
L
ZFlive
Actual real-time spectra.
How to Measure
Select the page K•
SET
with the rotary wheel
.
Measure the correction value k1 as described in the chapter •
Sound Level Meter: Correction Value K
SET. (k2 is not required
for SLV)
Press the start button •
.
The icon • LOG flashes during the ongoing measurement. In
the case where the profile #SLV 100dB & Audio has been
selected the icon AUD flashes, indicating the additional au-
dio wav-file logging.
During the measurement you can monitor the actual sound •
level L
Aeq5”
+k1 in order to remain below the permitted lim-
its; alternatively you may monitor the spectrum analyzer for
tracing possible feedback frequencies.
Press stop •
to finish the measurement.