Class 1 Sound Calibrator
(Refer IEC61672-1 paragraphs 5.2.1, 5.2.7, 9.3d)
If an obligatory calibration is required for the sound level meter,
in accordance with the type approval, then the Class 1 Sound
Calibrator CAL200 should be used.
Technical Details
Type: Larson Davis CAL200•
Level correction: -0.1 dB•
Calibration frequency: 1 kHz (= reference frequency)•
Calibration level: 114 dB (= reference sound pressure level)•
Details for Calibration
The calibration has to be executed according to the instructions
given in chapter “Calibration”. Ensure that the calibrator is set
to 114 dB.
Add the free-field correction values to the reference sound
pressure level, and enter the result into the XL2-TA under “Cali-
bration Level“.