User manual

190
Data Management
Assisted A pop up menu appears after each measure-
ment. You may select to keep or delete the
stored measurement records.
5
Save Test
Stores the measurement in the selected project folder.
6
Save Screenshot
Stores the XL2 screen in the selected project folder.
7
Load Test
Loads back previously-stored result data from the SD Card
for detailed examination on the instrument screen.
8
Append Mode
The append mode stores one or more results into the same
data file, simplifying data analysis and handling on the PC.
Application example:
When measuring the acoustic spectrum at various loca-
tions within the same venue, the append mode allows you
to store all the measurement results into one single data
file.
9
Free Memory
Displays the remaining available memory on the SD Card.
The XL2 prevents a memory overflow. SD Cards are not de-
signed for a 100% memory usage; the XL2 keeps at least
2% or 50 MB empty.
The XL2 deactivates the audio recording 50 MB before this
limit is reached and reports this on the display.
In the case where only 2% or 50 MB remains, the XL2
deactivates the data logging and reports this on the dis-•
play.
disables Save Test and Autosave•
saves the report at the end of the current measure-•
ment.