Data Management
12. Data Management
The XL2 stores all acquired measurement data including real-
time information onto the removable SD Card. Additionally,
wav-files might be recorded and individual voice notes added
for a complete documentation of the measurement.
Carry out a sound level measurement.•
Select the memory menu •
in the upper menu bar.
Press enter • . The memory menu opens.
1 2
4
3
5
6
7
8
9
10
1
Create New Project Folder
Add a new folder with an individual project name. The maxi-
mum length of the project name is 16 characters. A fast
way for creating new folder names is
Select an existing folder name with • DIR
Select • NEW
Rename the folder and confirm with start •
, thereby
creating a new folder name.
2
Select Project Folder
An overview of all existing project folders is shown. Select
a folder to store the results of the measurements.
3
Project Folder Name
All measurements are stored in this project folder.
4
Autosave
Setting of auto or manual data file naming and saving
On The results are named and stored automati-
cally in the selected project folder.
Off A question mark appears in the upper menu
bar
at the end of the measurement. You
can choose to individually name and store
each single measurement.