User manual
180
Capture + Tolerances
#FreqScale
Spacing
Defines the frequency scale spacing in the X-
axis for the tolerance data. (not mandatory)
lin Tolerance data displayed with
linear frequency spacing, e.g.
used in the FFT function. (de-
fault setting)
log Tolerance data displayed with
logarithmic frequency spacing
e.g. used in the 1/12 Oct + Tol
function.
#AutoStart Enables auto start function. (not mandatory)
True Auto start is active. The run indi-
cation displays A. The measure-
ment starts automatically after
the trigger level is reached.
Alternatively the measurement
may start manually by pressing
start
.
False No auto start function. (default
setting)
#AutoStartTrig-
gerLevel
Sets the sound level in dBZ for an automated
level trigger; the passed/failed measurement
starts as soon as the set level e.g. 95 dB is
reached or exceeded. The actual input level
can be monitored at the input range bar. (man-
datory if #AutoStart is True)
#AutoStartSet-
tlingTime
Defines the settling time, after the auto start
trigger level is exceeded, until the start of the
measurement; the time is set in 100 millisec-
ond steps e.g. setting 0.5 equals a settling
time of 0.5 seconds. If the input level drops
below the auto start trigger level during this
period, no measurement is carried out. (not
mandatory)
#MeasTime Defines the measurement time in 100 millisec-
ond steps; e.g. setting 1.5 equals a measure-
ment period of 1.5 seconds. (not mandatory)
#FailDeadTime Defines the delay for a failed condition in sec-
onds; for example the XL2 triggers “failed” if
the measurement result is beyond the toler-
ance limits for e.g. 60 seconds. This feature
supports condition monitoring applications.
(not mandatory)