User manual
175
Capture + Tolerances
Tolerance Menu
2
1
3
1
Capture & Start Tolerance
Opens capture and tolerance menu.
2
Tolerance Type
Select between high/low, high or low tolerance band based
on the reference data set.
3
Frequency Range
Setting of start and stop frequency of the tolerance band.
Advanced Start Functions
You may start the passed/failed measurement in one of the
following ways:
Press the XL2 start button •
.
Activate the automated level trigger in the tolerance file.•
Activate the digital input 1 via the external I/O interface; e.g. •
a foot switch or a PLC control.
Digital I/O Interface
Result Passed: Output 1
Result Failed: Output 3
Start Measurement: Input 1
Tolerance Management
Tolerance curves can either be imported from txt-files or direct-
ly derived from captured measurements. The XL2 calculates
tolerance bands based on
Single captured results•
txt-files generated manually on a PC•
Mean average of multiple captures•
Min/Max curves of multiple captures•