User manual
174
Capture + Tolerances
1
PASSED / FAILED Result
PASSED The actual measurement result is within the
tolerance band.
The actual measurement result is outside of
the tolerance band. Exceptions are visual-
ized in any out-of-tolerance frequency band.
2
Tolerance Bandwidth
Adjusts the tolerance bandwidth based on the captured ref-
erence spectrum in dB.
3
Level Offset
Provides a level offset in dB to shift the tolerance band up
or down.
Failed Result
The XL2 Analyzer compares spectral measurements against a
tolerance band and visualizes exceptions
1
in every frequen-
cy band.
1
Select another Tolerance File
Select the tolerance • Ref with the rotary wheel .
Confirm with enter •
.
Select the new tolerance file from the SD Card or the internal •