User manual
149
Vibration Meter
9
Page Selector X-Scale and Parameter Setting
1k7 Shows FFT result of the frequency band range.
23
Hz - 1.687 kHz in a resolution of
11.72 Hz with 143 bins shown on the display.
400 Shows FFT result of the frequency band range.
5 Hz - 421.02 Hz in a resolution of 2.92 Hz with
143 bins shown on the display.
100 Shows FFT result of the frequency band range.
1 Hz - 105.00 Hz in a resolution of 0.73 Hz with
143 bins shown on the display.
Usr User Range
(applicable with optional Extended Acoustic Pack
or Spectral Limits Option)
1 Hz - 20 kHz in zoom mode with a minimum reso-
lution of 0.366 Hz and 143 bins displayed.
Set Selection for FFT Windowing:
Hann• : Default setting
Dolph-Chebyshev• : for analyzing small signals
(e.g. harmonics) close to a dominant signal.
The page button
switches these display modes.
Oscilloscope
The scope function visualizes the waveform of the electrical in-
put signal. It automatically triggers to the fundamental frequen-
cy and selects the scaling of the X-axis (time) and Y-axis (level)
accordingly. The fundamental, or most dominant, frequency of
the input signals is displayed above the scope screen.
The scope scaling is fixed.